Coal slurry thickener feed and underflow: online solids concentration

Coal preparation

Online solids concentration for coal preparation thickeners — feed at 50–100 g/L in part-full gravity lines, underflow at 200–450 g/L — with a non-nuclear ultrasonic meter that has nothing in the flow for flocculated slimes to hang on.

Applicable industries
Coal slurry thickener feed and underflow: online solids concentration

PISONICS

PS7000 Series

Ultrasonic Slurry Density Meter

Acoustic impedance · non-nuclear · sensor flush with the pipe wall

Coal Preparation Thickeners — Feed and Underflow Solids Concentration

—— A dilute feed that often runs part-full, a dense underflow that is often pumped intermittently ——

【Key measurements: solids concentration g/L / slurry density】

Two Points, an Order of Magnitude Apart

In a coal preparation plant the thickener feed sets the flocculant dose and the thickener underflow sets how hard you can pull the bed and how the filter press is fed. Both are slime-water concentration measurements, but they sit an order of magnitude apart: the feed is dilute and usually runs part-full in a gravity line; the underflow is dense and pumped, often intermittently. Plants cover these points with nuclear gauges, with insertion fork meters, or with grab samples. This page explains why forks foul here, and what a non-nuclear ultrasonic meter needs to work on both lines.

PointTypical lineConcentrationApprox. % solids w/wDensity above water
① Thickener feedLarge-bore gravity line, often part-full50–100 g/L5–10 %19–45 kg/m³
② Thickener underflowFull bore, pumped, often intermittently200–450 g/L18–39 %75–200 kg/m³

Figures assume a slime solids density of 1.6–1.8 g/cm³. Across the whole feed range the slurry density moves by only about 20 kg/m³ — roughly 0.4 kg/m³ per g/L — while a 20 °C swing in water temperature moves water density by 4 kg/m³, the equivalent of about 10 g/L at the feed. The feed is a small-signal point: temperature compensation, sampling method and the acceptance criterion matter far more than they do on the underflow. The slurry concentration calculator converts between density, mass and volume concentration (g/L = mass fraction × slurry density).

Why a Fork Stops Reading True Here

A vibrating fork weighs the thin layer of liquid its tines drag along. Anything that sticks to the tines becomes permanent added mass. In coal slime that happens four ways:

  • Coating. Fine clay-rich slime and flocs build up at the root of the tines and between them. The reading drifts high and still looks plausible.
  • Bridging. Flocs or fibre bridge the gap between the tines; the reading jumps or the meter stops resonating.
  • Bubbles. Air from a drop-in feed or a mixing point clings to the tines and biases the reading low.
  • Wear. Coarse particles in the underflow thin the leading edge; the frequency drifts slowly.

This is a consequence of putting a resonator in the stream, not of any one brand. Hard coatings slow wear; they do nothing about build-up. We make a fork meter too (PS7400) — it belongs in clean liquids, chemical acids and alkalis, and tanks, not here. See tuning fork density meters for the full boundary.

How PS7000 Measures

PS7000 measures acoustic impedance with a single chirped ultrasonic transducer. The sapphire window sits flush with the pipe wall and does not project into the flow, so there is no root or gap for flocs to hang on. There is no radioactive source, so no licensing, leak testing or disposal.

Alongside the concentration it reports a 0–100 gas index, echo energy, signal quality and a fault word over Modbus; the two 4–20 mA loops follow NAMUR NE43 and go to 3.5 mA on a fault by default, so the DCS can tell a failure from a reading.

It has its own failure modes, and we state them: a mud film on the window makes the reading creep up and flatten out (echo energy and signal quality show it); free gas biases acoustic readings high, which only siting can avoid; and a change of coal calls for a calibration check.

Mounting Decides the Outcome

① Underflow (full bore)

  • Prefer a vertical upward run. On a horizontal line, mount at mid-height on the side of the pipe, away from settled solids at the invert and air at the crown.
  • Straight run ≥ 10D upstream and ≥ 5D downstream (5D / 3D minimum). Keep velocity above the deposition velocity — about 1.5–1.7 m/s for a DN200 (8-inch) line in coal slime.
  • On intermittent pumping, gate the measurement with the pump-running signal: after a stop the slurry settles and the reading creeps up without meaning anything.
  • Provide a flush connection, and set the 20 mA point well above the highest expected concentration.

② Feed (part-full)

A part-full line can be measured as long as the sensor face stays submerged under every operating condition. On large existing lines use the weld-in boss (Ø60 mm hole) on the lower half of the side wall, 30°–60° below the horizontal centreline: the lowest operating level limits the angle from above and any deposit at the invert limits it from below. For a DN600 (24-inch) line:

Angle below centrelineWindow centre above invertMinimum operating level needed
30°150 mm≥ 227 mm
45°88 mm≥ 160 mm
60°40 mm≥ 107 mm

Sensor positions on a part-full DN600 thickener feed line and a full-bore DN200 underflow line

The feed angle is chosen between the lowest operating level and the deposit at the invert (level and deposit in the figure are an example); on the underflow the sensor sits at mid-height on the side

  • Required level = top edge of the hole + 50 mm margin. Take the lowest level at low load and during start-up and shutdown, and check against the measured bore.
  • Put the sample point beside the sensor at the same height, so any vertical concentration gradient is built into the calibration.
  • Keep clear of drop-ins, side-stream impacts and flocculant mixing points upstream.
  • If the level falls below the window the reading drops to the bottom of the range with no fault code — gate it with feed flow or level in the DCS.

Calibration and Acceptance

  • Use oven-dried gravimetric samples as the reference, not quick density-cup checks; sample at the sensor at a recorded time, and take meter values from the DCS or cloud log.
  • Spread calibration points across the range (at least three near each end). The meter fits Y = aX + b with up to 20 points and, from three points on, refuses to write a fit with r² below 0.8; calibration is done online, without a shutdown.
  • Accept on at least 20 paired samples, judged on mean bias, trend agreement and the standard deviation of the differences. For the underflow the ±1 % (w/w) specification is a reasonable basis; for the feed, ±1 % w/w is about ±10–11 g/L — too coarse to be a useful criterion, so agree the figure from trial-period data.

Buying from China

CE marked; Ex d IIC T6 Gb available on order. Quotations in USD or EUR, EXW / FOB / CIF, with HS code and declaration of conformity. Send the line sizes and materials, the lowest operating level on the feed line, the underflow pumping pattern and the water temperature range, and the selection takes one exchange.

Selection support

Voices from users of this product

"Our original tuning fork and differential pressure meters on the absorber gypsum discharge main had recurring problems with bubbles and scaling — we had to shut down weekly to clean them. After switching to PS7000, both problems disappeared. Basically maintenance-free now, accuracy is stable, and it fully meets our FGD process control needs."

Thermal Control Foreman Wang
Thermal Control Specialist
A certain thermal power plant in Inner Mongolia

"After switching to the PS7000, our overflow density readings finally stabilized — we stopped tuning reagent dosing by feel. The unexpected win was not having to clean the sensor weekly; our previous radiometric meter needed window-wiping almost daily in the scaling slurry."

Director Li
Mineral Processing Workshop Director
A certain copper mining enterprise

"Our potash blending tank is a harsh environment — KCl near saturation, 30~40% crystal content, temperature swinging 5~20°C. Traditional density meters can't hold up here. After two weeks of PS7000 service, the deviation from manual lab samples stayed in the 0.5~0.8% range, even during concentration peaks. No anomalies."

Director Xie
Process Engineer
A potash fertilizer plant in Qinghai

FAQ

Can an ultrasonic density meter work on a part-full thickener feed line?

Yes, provided the sensor face stays submerged under every operating condition. That comes from the mounting height: the lower half of the side wall, below the lowest operating level and above any deposit at the invert. On a DN600 line with the sensor 45° below the centreline, the level must never fall below 160 mm.

When the level drops below the window or the line runs dry the reading is invalid and the meter raises no fault code, so gate it in the DCS with feed flow or level.

Won't the ultrasonic sensor foul the way a fork does?

A film can form, but the mechanism is different. The sapphire window is flush with the pipe wall, so there is no tine root or gap for flocs to lodge in. If the window does foul, the reading creeps up and flattens while echo energy and signal quality change with it — visible, not silent. Put the point on the routine inspection round and fit a flush connection.

Why does the underflow reading keep climbing after the pump stops? Is the meter faulty?

No. With the pump stopped the slurry settles in the line, solids gather at the sensor and the reading climbs slowly and unnaturally smoothly; none of it is a process value. Gate the measurement with the pump-running signal and hold the last good value while stopped. A genuine instrument fault drives the loop to a fault current instead (3.5 mA by default, per NAMUR NE43).

Why not accept the thickener feed meter on its ±1 % specification?

The specification gives ±1 % by mass concentration, which is one percentage point. On a 50–100 g/L feed that is about ±10–11 g/L, or 11–21 % of reading: neither a fair description of performance nor a useful acceptance figure, and narrowing the 4–20 mA span does not change the instrument's uncertainty.

Agree the feed criterion from the standard deviation of gravimetric comparisons made during the trial. On the underflow, at 200–450 g/L, ±1 % is about 3–6 % of reading and the specification is a reasonable basis.

How is dredge production (dry solids per hour) measured?

You cannot get it from density alone. Dry-solids production P_dry (t/h) = Q x Cv x rho_s, where Q is the volumetric flow (from a flow meter), Cv = (rho_m - rho_w)/(rho_s - rho_w) is the volumetric concentration from the mixture density rho_m (from an inline density meter such as the PS7000), and rho_s is the dry-solids density. The Pisonics Dredge Production Monitor reads the density and flow meters over Modbus and computes this every second, with dashboard and shift totals. See /guides/dredge-production-calculation-density-flow.

Can I measure dredge slurry density without a radioactive source?

Yes. An ultrasonic acoustic-impedance meter (PS7000) reads discharge-line slurry density inline (accuracy ±1 % by mass concentration, calibrated on the dredged material) with no radioactive source, so there is no shipboard radiation licence, no port-inspection delay and no radiation-safety officer. A clamp-on option (PS7010) allows no-hot-work retrofit, but it needs low gas content and is ruled out on lines that draw in air. See /guides/non-nuclear-density-meter-for-dredgers.

What density meter should go on a cutter-suction dredger discharge line?

A PS7000 ultrasonic acoustic-impedance meter suits CSD discharge lines (DN50-DN1000): flush sapphire window for abrasion, Chirp wideband to reject entrained air, non-nuclear. For no-cut retrofit use the clamp-on PS7010. Pair with a flow meter and the Dredge Production Monitor to also get dry-solids t/h. See /industries/dredging.

Is the PS7000 ultrasonic density meter a radiometric device? Does it need a radiation license?

The PS7000 is an acoustic-impedance ultrasonic density meter with no radioactive source whatsoever. No radiation license is required. It uses only piezoelectric transducers to send and receive ultrasonic signals — the same physical principle as medical and NDT ultrasound.

If you're currently using a Cs-137 / Co-60 source-based meter and want to remove the regulatory burden, PS7000 is a practical alternative; it installs differently, so check straight run and spool fit before a retrofit. A gamma gauge with an exempt-activity Na-22 source (< 1000 KBq) also needs no radiation license, but we no longer supply that class of instrument.

Can an ultrasonic concentration meter measure mine backfill slurry?

It can, depending on the slurry. The PS7000 acoustic-impedance meter covers 0–80 % by mass and its window sits flush with the pipe wall, so it suits fine-grained backfill slurry on a full pumped line (the datasheet recommends at least 65 % by mass finer than 75 μm). Paste close to the 80 % limit, slurry carrying coarse aggregate or coarse waste rock, and scaling cemented mixes need window wear, scaling and representativeness assessed first; the standard pressure rating is 1.6 MPa, and higher-pressure sections such as backfill pump discharges are rated case by case. It is a non-nuclear alternative to Cs-137 gauges — see /industries/mining.

Can PS7000 really measure stably in bubbly mining slurries?

Yes, with one distinction.

The PS7000 employs a linear frequency-modulated (Chirp) acoustic impedance algorithm: after transmitting a broadband ultrasonic pulse, the host unit analyzes the echo in the frequency domain, and multiple-reflection interference from fine, dispersed bubbles is identified and suppressed. That is the main difference from conventional single-frequency reflective meters. The meter also outputs a 0–100 gas index and a bubble flag, so the DCS can tell a real change in concentration from a bubble-driven high reading.

Heavy free gas (pump suction drawing air, a falling stream entraining it, a point directly above an aeration header) or a continuous gas film over the sensor face pushes the reading high or stops it altogether. That is a limit of acoustic measurement: it is avoided by where the sensor goes, not corrected in calibration.

At the gypsum discharge line of an absorption tower in a thermal power plant in Inner Mongolia (under conditions of continuous air oxidation that generate dense bubbles), the PS7000 has been operating stably for several years after replacing the original tuning fork concentration meter.

What installation requirements does the PS7000 have?

PS7000 installs in one of two ways, and both have the same process requirements:

  1. Spool type: a flanged measuring spool (ANSI / DIN / JIS) goes into the line, with the sensor face made flush with the bore at the factory. Weld-in type: a Ø60 mm hole and a welded boss on the existing pipe, with no cut-out, which suits large existing mains;
  2. Straight run ≥ 10D upstream and ≥ 5D downstream (5D / 3D minimum where space is tight), away from pump discharges, valves and bends;
  3. Prefer a vertical run with upward flow. On a full horizontal line, mount at mid-height on the side of the pipe, clear of settled solids at the invert and air at the crown;
  4. The pipe need not run full, but the sensor face must stay submerged under every operating condition — on a part-full line, mount on the lower half of the side wall, below the lowest operating level and above any deposit at the invert;
  5. Keep velocity above the deposition velocity and preferably no more than 5 m/s. Readings taken while the line stands after a pump stop do not represent the process; gate them with the pump-running signal in the DCS;
  6. Line sizes DN50 to DN1000 (larger on request). The window is sapphire; for heavy abrasion choose 316L with special ceramics or a 2205 duplex probe, and for strong corrosion a PTFE lining.

For how to choose the angle and level on a part-full line, see coal slurry thickener feed and underflow. If the line cannot be drilled, hot-worked or shut down, select the PS7010 clamp-on meter instead.

PS7000 vs nuclear density gauges: which costs less over the life cycle?

On purchase price alone, ultrasonic and nuclear gauges sit in a similar bracket. The gap opens over 5 to 10 years of ownership.

Hidden cost list of a Cs-137 / Co-60 nuclear gauge:

  • Radiation safety licensing and annual reviews, plus operator training and certification;
  • Licensed transport and installation filing for the source;
  • Dose monitoring and record keeping during service;
  • Source replacement as activity decays (purchase, transport, commissioning, return of the old source);
  • End-of-life disposal of the spent source — often the single largest bill.

PS7000 acoustic-impedance ultrasonic gauge: no radioactive source and no permits of any kind; non-intrusive sensor that does not project into the flow, so no insertion-type erosion or build-up points, sensor life of 5 years or more, virtually maintenance free with no consumables. Power plant, potash and iron ore sites have run 2+ years at near zero maintenance.

Bottom line: on a 5-year basis the total cost of ownership of the PS7000 is typically far below a nuclear gauge. Where a nuclear principle is genuinely required, a gamma gauge with an exempt-activity Na-22 source is genuinely a lighter route than Cs-137 and needs no license, though the roughly 2.6-year half-life still implies periodic source renewal. We no longer supply that class of instrument; it is noted here for completeness.