Inline concentration monitoring for juice, purée and concentrate

Food · Beverage

Inline refractometric concentration for sucrose, fructose, juice, purée, jam and stock: unaffected by colour or pulp turbidity, for evaporation end-point and blend ratio control, with CIP / SIP and a sanitary adapter.

Applicable industries
Inline concentration monitoring for juice, purée and concentrate

PISONICS

PS7110 Series

Inline Process Refractometer

Critical angle · thin-layer interface · indifferent to colour and turbidity

Inline concentration monitoring for juice, purée and concentrate

— Sucrose · fructose · juice · purée · jam · stock —

Measured: Soluble solids (Brix) / refractive index nD

Inline concentration monitoring for juice, purée and concentrate

Process background

On a juice or purée line, concentration is both a quality specification — soluble solids, which the finished product has to meet — and a process variable: when to stop evaporating, and in what ratio to blend. It is also one of the harder liquids to measure optically, carrying pulp, fibre, pigment and air all at once.

What critical-angle refractometry brings here is that it does not need the light to get through. Light strikes the interface between prism and liquid, totally reflects beyond the critical angle, and a linear CCD reads where the shadow line falls. Pulp and pigment sit outside that thin layer; the optical path never reaches them. Dark concentrate and pulpy purée read the same as clear juice.

What the method can and cannot do here

What it can do:

  • Soluble solids as Brix — evaporation end point, blend ratio, finished-product conformity
  • Dark, highly turbid liquids carrying pulp and air — none of it affects the reading
  • Continuous tracking — a few laboratory samples a day cannot show what happened in between

What it cannot — stated here so nobody specifies it into the wrong duty:

  • Total solids including insoluble pulp — refractometry counts only the dissolved fraction
  • Telling fructose from sucrose — both raise the index, and the reading is the total
  • Staying accurate through pectin filming or sugar deposit — high-viscosity purée should carry a CL1 from the start

Selecting the PS7110

VariantPrismProcess temp.RI rangeSuits
PS7110-AHigh-strength optical glass0 – 60 °C1.33299 – 1.51782Ambient, non-abrasive, span within 0–90 %Brix
PS7110-BSapphire0 – 70 °C1.33299 – 1.57041The default for most inline duties; 0–100 %Brix
PS7110-CSapphire−10 – 100 °C1.33299 – 1.57041Evaporation, and points before and after pasteurisation

The connection is sanitary: the PS7110-MT4 adapter, CIP and SIP compatible, cleaning temperature −30 to 120 °C. Where the face has to be inspected, the PS7110-MT2 / MT3 tee and cross carry sight glasses.

An evaporator outlet goes to the C variant (−10 to 100 °C); blending tanks and pre-fill points are fine on the A or B. Purée is viscous and coats, so specify the PS7110-CL1 ultrasonic cleaning unit by default — retrofitting one after the reading has drifted costs considerably more than fitting it first.

General specification:

ItemSpecification
Measured quantityRefractive index nD, converted to concentration through a scale (%Brix or wt% / vol%)
AccuracyConcentration ±0.1 %; refractive index ±0.0001; temperature ±0.5 °C
ResolutionConcentration 0.01 %; refractive index 0.00001; temperature 0.1 °C
Pressure≤ 1.5 MPa at the prism face
Cleaning temperature−30 – 120 °C; CIP and SIP capable
Wetted materialsSS316L plus prism; B / C accept PTFE / Hastelloy / titanium / tantalum
Outputs4–20 mA / RS485 / RS232 / USB
Hazardous areaEx ia IIC T6 Ga (intrinsically safe)

The scale has to be built for your medium

This is the most commonly misunderstood thing about refractometry, so it goes near the front.

What is calibrated at the factory is refractive index and temperature, unit by unit in standard units. That part is independent of the medium, so several instruments agree with one another. The concentration scale, however, belongs to the medium — the curve from index to concentration is a property of one liquid, and a scale built for one medium cannot be used for another.

In practice that is three things. One: tell us the medium and you get the matching scale. Two: a change of recipe, product or process stage means a change of scale, but not a trip back to the factory — RI verification, self-service modelling from sampled laboratory values and multi-point temperature compensation are all done on site. Three: be careful with multi-component systems — the same refractive index does not mean the same concentration. Two components that move the index the same way cannot be separated by refractometry, and that duty belongs to the PS7100 spectroscopic route.

Installation and upkeep

Refractometry reads a thin layer of liquid at the prism face, so a trustworthy reading rests on two things only: whether the face is always washed by liquid that represents the process, and whether it is clean.

  • The face must never point downward — a downward face collects deposit, and deposit raises no fault; it shows up only as a slow one-way drift
  • In a vertical line the flow must run upward — a downward run drains when the pump stops, the prism is left uncovered, and the reading stops meaning anything
  • On a horizontal line, entry from below (face up) or from the side (face vertical) are both fine; entry from the top is not permitted
  • Four mounting schemes: valved bypass (recommended — service it without stopping the main line) / direct in-line (cheapest, service needs a shutdown) / pumped tank loop (steadiest conditions) / tank wall (simplest)
  • 2″ and under takes the PS7110-MT1 small-bore adapter; where the face has to be inspected visually, the MT2 / MT3 tee and cross carry sight glasses
  • Media that can carry grit or abrasive particles must be filtered upstream — the detection window is a wear part

The cleaning strategy belongs in the design, not in a response to a reading that has already drifted: sanitary processes use CIP and SIP directly; media that coat or film take the PS7110-CL1 ultrasonic unit for continuous cleaning; media that crystallise or deposit particles take the PS7110-CL2 high-pressure flushing unit on a timed cycle.

The instrument carries prism condition diagnostics: the position of the shadow line is the measurement, and its sharpness and contrast are a second quantity the same detector sees at the same time. A deposit blunts the edge before it shifts the position, so the instrument reports its condition before the reading is visibly wrong rather than quietly returning a plausible number. It does not replace looking at the face, though — putting that into the routine inspection is worth more than any coefficient correction applied afterwards.

Before you order

These are what decide the variant, the accessories and the scale. Sending them over is quicker than guessing from a datasheet.

  • Product and stage (single-strength / concentrate / purée / jam / blend) — this sets the scale
  • Brix working band and finished-product tolerance
  • Process temperature, particularly the real maximum at the evaporator outlet
  • Viscosity and pulp content — this decides whether a CL1 is mandatory
  • Connection type and pipe size
  • Whether the scale should be modelled on site from laboratory data

This page is a method and selection note, not a case study. The PS7110 has no publicly citable installed base in this industry yet, so there are no unit counts, saving percentages or payback periods here — we do not invent those. Give us the process conditions and you get a selection opinion you can check, which is worth more than a handsome case study.

In closing

The Pisonics PS7 range covers several measuring principles: acoustic impedance (PS7000), acoustic attenuation (PS7010), sound velocity (PS7020), spectroscopy (PS7100), refractometry (PS7110), Coriolis (PS7200), differential pressure (PS7300), tuning fork (PS7400) and microwave (PS7600). Different principles see different things, and selection always starts with what the duty actually needs measured.

The PS7110 series takes the critical-angle route: the measurement happens in a thin layer at the prism face, which is why colour, turbidity and bubbles do not interfere — and why the prism face has to be kept clean. Juice and purée are the textbook case of dark, turbid liquids whose control variable is nonetheless dissolved-solids concentration. Method and process line up.

For a non-standard span, special wetted materials, a hazardous-area version, or to have us build the concentration scale from your own sampled laboratory data, please get in touch with our engineering team.

FAQ

Does the PS7110 prism need cleaning, and how often should it be checked?

A clean prism face is the single precondition for a trustworthy reading. Coating, scale, crystals or abrasion change the reading directly, and the instrument does NOT raise a fault when they do — it shows up as a slow one-way drift on an unusually smooth curve. So the cleaning strategy belongs in the design, not in a response to a reading that has already drifted. The prism material depends on the variant: high-strength optical glass on the A, sapphire on the B and C. For media that coat or crystallise, the PS7110-CL1 ultrasonic cleaning unit keeps the face clean continuously and the PS7110-CL2 high-pressure flushing unit flushes it periodically; sanitary processes can use CIP and SIP directly, with a cleaning temperature range of −30 to 120 °C. Where the medium can carry grit or abrasive particles, filter them out upstream or the detection window will be worn away. Putting "look at the face" into the routine inspection is worth more than any coefficient correction applied afterwards; the sensor status diagnostics help, but they do not replace looking.