Industrial Online Density Meter Principles: 7 Technologies Compared

Every industrial online density measurement principle in one table — what each physically measures, where it wins, and the condition that rules it out.

There are seven principles in common industrial use for online density and concentration. None is universal. This page is the map; the six-step selection guide is the route.

The comparison

PrinciplePhysically measuresWins whenRuled out by
Ultrasonic — acoustic impedance Reflection at a boundary, plus sound speed Abrasive slurry, large bore, gas present, no source wanted Media where solids and liquid have almost no impedance contrast
Ultrasonic — attenuation Energy lost across a transmission path Cannot tap the pipe; retrofit and trial installs Entrained gas; bore above DN500
Ultrasonic — time of flight Speed of sound Clean binary solutions with a monotonic curve Slurry; any medium where the curve turns over
Tuning fork Resonant frequency of an immersed fork Tanks and lines, wide chemical range, good value Heavy abrasion; media that build up on the fork
Coriolis Tube resonant frequency, plus phase shift for mass flow Clean liquids needing density and mass flow together Two-phase flow; abrasive slurry; large bore economics
Optical Refractive index, or absorbance spectrum Clean liquids; chemical specificity; lab traceability Opaque abrasive slurry
Differential pressure Hydrostatic head over a known height Simple, inexpensive, well-understood Variable velocity, moving installations, gas in the leg
Microwave Phase shift through the medium High solids where water content dominates Media whose conductivity swamps the permittivity signal
Gamma / radiometric Attenuation of gamma photons Nothing may touch the process; extreme duty Licensing, transport and disposal obligations

Three questions that eliminate most of the table

  1. Is there abrasive suspended solid? If yes, Coriolis and optical are out, and tuning fork needs a wear-resistant tip.
  2. Is gas present, or does the process create it? If yes, transmission ultrasonic and Coriolis suffer most, and you should ask specifically how the candidate instrument behaves — and whether it tells you.
  3. Can the pipe be tapped? If no, the feasible set narrows to clamp-on ultrasonic, externally mounted gamma, or a bypass loop.

Working through those three before comparing accuracy figures saves most of the effort, because accuracy is rarely the binding constraint — survival in the duty and behaviour under upset usually are.

Accuracy is not comparable across principles

A specification of ±0.001 g/cm³ on a clean-liquid instrument and ±1 % FS on a slurry instrument are not on the same scale, and neither number tells you what the instrument will do on your process. Ask instead: what was the reference method, what medium was it calibrated on, and what happens when the process moves away from those conditions. A calibration certificate on the actual process liquid is worth more than a smaller number on a datasheet.