Spent emulsions and oily water — phase detection after demulsification

Environmental · Water treatment

Skim and decant decisions after breaking spent cutting fluid, rolling emulsion or oily wastewater. Here the rag layer is the normal state rather than an upset, so the amplitude channel earns its keep: the share of each batch spent in the emulsion state measures how well the break worked, a batch earlier than an oil-in-water lab result. The velocity margin depends on temperature and salt — thin at ambient on light oil, over 200 m/s once hot and salted.

Applicable industries
Spent emulsions and oily water — phase detection after demulsification

PISONICS | 西安派声

PS7021 series

Inline Ultrasonic Phase Separation Detector

Sound velocity + amplitude · integral spool

Spent emulsions and oily water — phase detection after demulsification

—— On this duty the amplitude channel matters more than the velocity ——

【Discriminators: emulsion share, plus the two velocity bands】

The Duty

Spent cutting fluid, rolling emulsion and oily wastewater are all treated the same way: acid, salt, reagent or heat to break the emulsion → settle → skim oil, decant water. The questions are whether the skim line is running oil or still running rag, and whether the decant line is carrying oil out with the water.

What separates this duty from the others: the rag layer is the normal state, not an upset. Demulsification is never complete; the middle layer is always there, only thicker or thinner. The useful output here is less "which phase" than "how much rag is left" — which is the amplitude channel.

The Gap Depends on Conditions

This is the duty on this site that most needs a measured sample, and the reason is worth stating plainly:

ConditionOil phaseAqueous phaseGap
Ambient, light mineral oil, low-salt waterabout 1431–1477 m/s (same-order reference)1497 m/s20–66 m/s, close to the floor
As demulsified: 60 °C, acid and salt addedabout 1335 m/s1551 m/s and up with saltabout 216 m/s or more, comfortable

So the siting rule is blunt: measure on the hot line after demulsification, not at the outlet of an ambient holding pit. Heat and salt are things the process does anyway, and they are what opens the margin — the oil coefficient is negative (−2.75 to −3.6 m/s per °C), water's is positive, and salt adds roughly +11 m/s per wt% NaCl.

The oil column uses published olive and castor oil values as a same-order reference; the water column is the Marczak equation. A spent emulsion's oil phase (mineral oil plus esters plus additives) can only be measured, not looked up — the table is here to show which way temperature and salt move the margin.

Using the Amplitude Channel

  • Skim line: only an amplitude back at the clear-phase baseline means you are skimming oil rather than rag; a persistently low amplitude says the break is incomplete — adjust reagent or settle longer instead of skimming harder
  • Decant line: amplitude down while velocity stays in the aqueous band → emulsified oil in the water; recycle, do not discharge
  • As a trend: the share of each batch spent in the emulsion state is a direct measure of how well the break worked — a full batch earlier than an oil-in-water lab result
  • The default threshold is 6 dB below the clear-phase baseline; on this duty it usually gets retuned to the site's own oil

Maintenance: This Line Will Foul

  • Oil and reagent film on the window is certain, and the amplitude baseline follows it down — use that as the trigger for planned cleaning
  • Sapphire is dense and smooth; solvent or hot caustic restores it without pulling the sensor
  • Avoid points carrying heavy suspended solids (swarf, sludge): scattering holds the amplitude below threshold permanently

Where It Does Not Work

  • An emulsion that has not been broken: the whole line is rag and there are no two phases to tell apart — that is a reagent and process problem
  • Sludge: high solids, high attenuation; solids measurement is PS7000
  • Ambient, low-salt, light-oil combinations: the gap can fall below 30 m/s, so a measured sample comes before any proposal

To quote this duty we need: the waste stream (cutting fluid, rolling emulsion, oily water), how and at what temperature it is broken, salt and pH of the aqueous phase, roughly what the oil phase is, line size and material, suspended solids, and the measuring point. Send 500 mL of each phase as demulsified and we will measure both bands at your break temperature — on this duty we do not quote from handbook values.

FAQ

Dichloromethane is far denser than water — why not use a density or acoustic-impedance meter?

Because an ultrasonic density meter reads acoustic impedance, Z = ρ × c, not density, and dichloromethane happens to be dense and acoustically slow, so the two factors cancel. At 20 °C water is 1.48 MRayl and DCM about 1.45 MRayl, and between roughly 15 and 20 °C they coincide — which is where most plants separate. Inferred against water's velocity, DCM reads about 0.98 g/cm³: lighter than water, the wrong way round.

The same pair differs by 390 m/s in sound velocity (1482 against 1090 m/s, about 26 %), with opposite temperature coefficients, so warming widens the gap. That is why this duty takes the sound-velocity model (PS7021) rather than the impedance one (PS7000). DP and Coriolis meters do read true density and can separate the pair in principle, but they cannot keep up with a draining batch.

How is the rag layer handled — can it be mistaken for the aqueous phase and let through?

The rag layer is a third state, not a grey zone between two. Two quantities catch it.

First, velocity: a dispersion falls between the two bands (Wood's equation, first order, puts 30 vol% DCM near 1338 m/s and 50 vol% near 1256 m/s), and anything outside the clear-phase bands is not reported as a clear phase. Second, and more decisively, amplitude: droplet scattering weakens the echo, and 6 dB below the clear-phase baseline is reported as emulsion / not-decidable by default.

The interlock direction is fixed: on emulsion the instrument holds the valve and raises an alarm rather than switching. It will stop and wait for an operator before it lets a rag layer pass as either phase. The emulsion itself is a separation problem; what the instrument can do is refuse to wave it through.

Our line is DN32 and PTFE-lined — can this be fitted, and what happens at the liner?

Yes. DN32 is one of the standard bores (DN25 / DN32 / DN50). The lined build is a steel shell with a PTFE/PFA liner and sapphire windows that pass through the liner and wet the process.

Measuring through the liner is not an option: PTFE is 2.97 MRayl and strongly attenuating, and the bond between liner and shell changes with temperature and permeation — that turns a fixed acoustic path into a drifting one.

Two ordinary lined-pipe rules also apply. Wetted seals are PTFE or FFKM; ordinary FKM rates only fair against methylene chloride. And the shell's vent holes must stay clear — chlorinated solvents permeate fluoropolymer liners slowly, and trapped permeate lifts the liner off the wall.

Against conductivity and photoelectric (sight-glass) detectors — when should each be chosen?

Choose by whether the premise holds; each method rests on one.

Conductivity assumes the aqueous phase conducts, the organic does not, and an electrode may wet the process. Where that holds it is the cheapest answer. Pure-water washes, wet organics and lined pipe that is awkward to tap all soften it.

Photoelectric / sight glass assumes an optical difference. Turbidity- and colour-based units fail outright on dichloromethane and water — both are clear and colourless — while a refractive-index unit does work (1.424 against 1.333). Both share one weakness: a fouled window.

Sound velocity (PS7021) rests on two conditions only: the phases differ by at least 30 m/s, and the pipe runs full. It is independent of colour, turbidity, conductivity and density, which is why it holds where both phases are clear, both are non-conducting, or the aqueous conductivity moves. The price is a two-phase calibration on site, repeated at product changeover — five minutes, but somebody has to do it.

The item-by-item comparison is here.

Does commissioning need a shutdown, and how are the samples taken?

No shutdown. Calibration needs two velocity bands, and both samples come out of a normal batch: one while the light phase is running steady, one while the heavy phase is, 200 mL or more each. The instrument reads them itself — nothing goes to a lab. It usually fits inside one batch.

Three cautions: sample at process temperature (velocity is temperature sensitive, organics run −3 to −4 m/s per °C); draw the sample near the measuring point; and take the aqueous phase at its real salinity — clean water is not a substitute.

What you end up with is a velocity-versus-temperature table for both phases, delivered with the instrument. If the process temperature changes later, the thresholds shift along that table without a service visit.

How do the three relays drive a diverter valve, and which way does it fail?

Three volt-free contacts carry heavy, emulsion and light. The usual wiring: heavy closed → diverter to the heavy receiver; light closed → light receiver; emulsion closed → hold the current position and alarm for an operator to confirm.

The failure direction is the first thing to settle. Gas, flashing and window film all push the instrument toward not-decidable rather than toward the other phase, so making not-decidable mean stay put runs with the grain of the disturbances. Defaulting it to either phase turns every disturbance into a carry-over incident.

Keep the manual mode and the existing sight glass: the instrument supplies the judgement, not the valve. 4-20 mA trends velocity in the DCS; RS485 exposes phase, velocity, amplitude and temperature for the batch record.

How is it accepted on site — what do you compare it against?

The output is a phase decision, not a concentration you can compare against a titration, so acceptance looks different from a concentration meter's. Three workable tests:

1. Timing against the operator. Run a number of batches and align the instrument's switch instant with the operator's call at the sight glass. The site sets the tolerance; what usually matters is whether the instrument is ever late, since late means carry-over.

2. Sampling around the cut. Draw a sample either side of the instrument's switch instant and check that carry-over sits inside process limits.

3. The velocity itself. Measured band values should match the calibration table, and batch-to-batch drift in one phase should stay within the repeatability seen at calibration (±0.2 m/s). This is the test that separates "the instrument drifted" from "the material changed".

Can it measure how thick the rag layer is, or where the interface sits in a tank?

No — those are different instruments. PS7021 is an inline spool and knows only what is crossing its acoustic path right now: heavy, emulsion or light.

Interface height in a tank or settler belongs to an interface level instrument (magnetostrictive, radar, ultrasonic level, capacitance), mounted on the vessel rather than in the line. Rag layer thickness is the same question.

There is a useful indirect answer, though: time the instrument spends in the emulsion state during a drain, multiplied by flow, gives the volume of rag in that batch. It is not an interface height, but as a trend for how well the break worked or whether settling time is enough, it is often the more direct number.

Does it need a straight run? Can it go right after a pump or a control valve?

No straight run is required. The path length is fixed by the spool and the discriminators are velocity and amplitude, which do not care about the flow profile — unlike DP, vortex or magnetic meters, which live on it.

The position still matters, but for a different reason: phase and gas. Pump discharges, the downstream side of control valves and other throttling points generate gas or flashing continuously, which holds the amplitude down and degrades three-state discrimination. Mount upstream of them, or further downstream where the flow has settled.

Three other rules: a full line is not negotiable; on horizontal runs put the path across the horizontal (3 and 9 o'clock), away from gas at the top and solids at the bottom; and keep the measuring point close to the diverter, because that hold-up is the mixed volume every batch pays.

What about high temperature, sanitary execution and CIP?

Temperature: −20 to +80 °C standard, −20 to +120 °C for the high-temperature variant. Heat actually helps the discrimination: organic velocity falls with temperature while water rises below 74 °C, so the bands spread. Edible oil washing at 80-90 °C is the textbook case.

Sanitary: the standard spool is industrial and flanged. Tri-clamp, polished bore and drainable geometry are a custom build with their own lead time and price — say so at selection rather than assuming the standard spool can enter a sanitary zone.

CIP: the sapphire window is dense and smooth, so hot caustic, hot water or solvent all clean it in place. Compare the amplitude baseline before and after and you know whether the clean worked — far easier than scheduled teardown.

Can it still work if the line carries solids or crystals?

It depends on the load. A light suspension does not move the velocity reading but scatters sound and lowers the echo amplitude — and amplitude is what catches the rag layer, so with heavy solids three-state discrimination degrades to two: light and heavy still separate, emulsion and "solids present" no longer do.

In practice: put the point after settling, filtration or a steady pump; at commissioning re-baseline the amplitude against the real solids-bearing fluid rather than against clear liquid; and if solids content is itself the quantity you want, that is another instrument — PS7000 and the acoustic-impedance method.

Crystals add one more wrinkle: they grow a film on the window. The film shows in the amplitude baseline first, which makes that baseline a useful clean-me reminder.

How wide does the velocity gap have to be, and can it be estimated first?

The floor is 30 m/s at working temperature. Common organic-against-water pairs are far above it: carbon tetrachloride is 572 m/s from water, dichloromethane 428, ethyl acetate 413, toluene 170, m-xylene 155 (handbook values at 25 °C). For those the gap is never the constraint.

Three cases do need arithmetic first: oil against oil (diesel and kerosene are 74 m/s apart), oil against water at ambient (edible oil and water, about 66), and formulated fluids — peroxide working solution, SX organic, a demulsified waste oil.

How to estimate: take handbook values at 25 °C, carry both to working temperature with their coefficients (organics −2.7 to −4.9 m/s per °C, water about +2.4 below 74 °C), then add salt on the aqueous side (roughly +11 m/s per wt% NaCl). Above 100 m/s you are safe; 30 to 100 needs a measurement; below 30, don't.

How many products can one instrument cover, and does a changeover need recalibration?

Eight recipes as standard, each holding a pair of velocity bands and its amplitude threshold; the list extends. Switch from the panel, or let the DCS write a Modbus register — the second follows the batch instruction and removes the "nobody switched it" class of incident.

A product that has never been calibrated needs one pass: two samples, two readings, one stored pair — five minutes, no return-to-factory sampling. Coming back to a known product just recalls its recipe.

When to redo an existing recipe: the composition of that phase moved (water content, salt, product concentration), the temperature band shifted, or acceptance found measured velocity off the stored band by more than the repeatability. Every recalibration records its time and both velocities for the batch record.

How often does the sapphire window need cleaning, and how do you know?

No fixed interval — it depends on the duty. What actually helps: the instrument tells you.

As film, crystal or oil builds on the window, the velocity reading does not move at first, but the amplitude baseline drifts down. Trend the clear-phase baseline: a slow decline means clean it; a sharp drop with velocity still in band is an emulsion slug passing, not deposit. The two shapes are easy to tell apart on a trend.

How to clean: sapphire is dense and smooth, so hot caustic, hot water or a normal solvent will do it, usually in place. On duties that will certainly film — edible oil, spent emulsions — put it on the CIP schedule. Afterwards check whether the amplitude baseline came back: that is the acceptance test for the clean.