PISONICS | 西安派声
PS7021 series
Inline Ultrasonic Phase Separation Detector
Sound velocity + amplitude · integral spool
Hydrometallurgy SX — organic / aqueous phase detection at the settler outlet
—— Organic into the tankhouse is the expensive mistake ——
【Discriminator: organic carry-over vs aqueous carry-over】
The Duty
Copper, nickel, cobalt and lithium hydrometallurgy all run leach → solvent extraction → stripping → electrowinning. The SX stage is a mixer-settler: the mixer emulsifies organic (diluent plus extractant) with aqueous, the settler lets them separate, organic overflows back and aqueous leaves. The question is which phase is in the outlet line right now.
Carry-over is asymmetric here too:
| Direction | Consequence |
|---|---|
| Organic into electrowinning | The expensive one — organic collects on the cell surface, a fire risk around hot anodes, and it shows up as organic burn on the cathode: contaminated plate, rework |
| Aqueous into the organic circuit | Acid and chloride carried into stripping and washing; extractant loading and degradation both accelerate |
Is the Gap Wide Enough
The organic phase is mostly kerosene-type diluent (handbook value at 25 °C 1324 m/s, coefficient −3.6 m/s per °C); with P204 / P507 or an oxime extractant it still sits in the 1300 to 1380 m/s range. The aqueous phase is a sulphate or chloride liquor: water is 1497 m/s at 25 °C and salt and acid raise it — roughly +11 m/s per wt% NaCl. The bands normally sit 180 to 250 m/s apart.
The exact pair still comes from site samples: extractant ratio, acidity and metal loading all move them, and they move with the circuit. That is why the thresholds are set at commissioning rather than in the factory.
What It Answers, and What It Does Not
| Question | This instrument | What to use instead |
|---|---|---|
| Which phase is passing now | Its job | PS7021, three states, interlocked |
| Where the settler interface sits | No | An interface level instrument, measuring height in the vessel |
| How many ppm of organic are entrained | No | Sampling; online, turbidity or optical |
| Organic-to-aqueous ratio | No | PS7110 / PS7600 |
| Metal concentration in the liquor | No | PS7020 sound-velocity concentration meter |
Where It Goes
- Settler organic overflow — watching for aqueous in the organic circuit
- Aqueous outlet / raffinate pump — watching for organic leaving with the aqueous; this is the one that protects the tankhouse
- Strip and wash outlets — same two-phase question, same discriminator
- Solids are normal here (liquor carries fines, crystals): a high solids load holds the amplitude down and three-state discrimination degrades to two, so pick a point after settling, on steady pump flow
To quote this duty we need: diluent and extractant types and ratio, aqueous composition (acidity, main salts, metal loading), operating temperature, line size and material, solids content, and the distance to the valve. Send 500 mL of each phase and both bands come from one measurement.