Product pipelines and solvent transfer — batch interfaces and water slugs

Oil and gas · Terminals

Two kinds of interface on one line, an order of magnitude apart. Oil against water is about 247 m/s — a solid water-slug call. Diesel against kerosene is 74 m/s: workable, but every pair has to be measured and compensation kept tight, since one degree of error eats 5 % of the margin. The page states what it does and does not do (no custody transfer, no interface volume) and how it sits beside density-based interface detection.

Applicable industries
Product pipelines and solvent transfer — batch interfaces and water slugs

PISONICS | 西安派声

PS7021 series

Inline Ultrasonic Phase Separation Detector

Sound velocity + amplitude · integral spool

Product pipelines and solvent transfer — batch interfaces and water slugs

—— Oil against water is easy, oil against oil is tight — sort out which question first ——

【Oil / water 247 m/s · diesel / kerosene 74 m/s】

Two Different Questions

One line carries two kinds of interface problem, an order of magnitude apart in difficulty. Separate them before anything else:

QuestionVelocity gap (handbook, 25 °C)What this instrument does
A water slug in the line
after pigging, a shutdown, or tank draining
about 247 m/s
diesel 1250 / water 1497
Solid. Two hundred-odd m/s apart leaves room for thresholds and hysteresis
A batch interface
diesel → kerosene, gasoline → diesel
about 74 m/s
diesel 1250 / kerosene 1324
Workable, but each pair must be measured and compensation must be tight: at −3.6 m/s per °C, one degree of error eats 5 % of the margin

Why Temperature Compensation Becomes Critical

On the earlier duties the phases sit two to three hundred m/s apart and a few m/s of thermal drift is noise. At seventy-odd m/s it is not:

  • Both products have nearly the same coefficient (−3.5 to −4 m/s per °C), so temperature does not spread the bands the way it does for oil and water — it translates them together
  • The built-in Pt1000 reads at the same place as the acoustic path, which decides the result here: the further the temperature point from the path, the more the compensation is a guess
  • Seasonal swing (a dozen degrees between winter and summer line temperature) belongs in the threshold table, not in a single ambient band

What It Does Not Do

  • No custody transfer. That is density and flow — PS7400 tuning fork and PS7200 Coriolis
  • No interface volume. It gives the moment the interface arrives, not how long the mixed slug is
  • No grade-to-grade calls within one product (two diesel grades, say). Those can be a dozen m/s apart, under the floor; scheduling records and a density meter do that job

Against Density-Based Interface Detection

Density is the established method here and it works — but on oil against oil it is just as tight: diesel and kerosene differ by about 0.04 g/cm³, near 5 % relative, the same order as the 5.6 % relative velocity gap. Both routes need site measurement; neither is exempt. The real decision is installation: a station that already runs an inline densitometer has no reason to add a second instrument for interfaces, while a point that needs both water slugs and batch interfaces, or where an insertion fitting is awkward, is easier with a spool.

To quote this duty we need: the product or solvent list in switching order, the temperature range of each, line size and pressure class, area classification, where the measuring point sits (inlet, outlet, ahead of tankage) and whether an inline density meter is already installed. Send 500 mL of each medium and we will measure a table at your line temperature.

FAQ

Dichloromethane is far denser than water — why not use a density or acoustic-impedance meter?

Because an ultrasonic density meter reads acoustic impedance, Z = ρ × c, not density, and dichloromethane happens to be dense and acoustically slow, so the two factors cancel. At 20 °C water is 1.48 MRayl and DCM about 1.45 MRayl, and between roughly 15 and 20 °C they coincide — which is where most plants separate. Inferred against water's velocity, DCM reads about 0.98 g/cm³: lighter than water, the wrong way round.

The same pair differs by 390 m/s in sound velocity (1482 against 1090 m/s, about 26 %), with opposite temperature coefficients, so warming widens the gap. That is why this duty takes the sound-velocity model (PS7021) rather than the impedance one (PS7000). DP and Coriolis meters do read true density and can separate the pair in principle, but they cannot keep up with a draining batch.

How is the rag layer handled — can it be mistaken for the aqueous phase and let through?

The rag layer is a third state, not a grey zone between two. Two quantities catch it.

First, velocity: a dispersion falls between the two bands (Wood's equation, first order, puts 30 vol% DCM near 1338 m/s and 50 vol% near 1256 m/s), and anything outside the clear-phase bands is not reported as a clear phase. Second, and more decisively, amplitude: droplet scattering weakens the echo, and 6 dB below the clear-phase baseline is reported as emulsion / not-decidable by default.

The interlock direction is fixed: on emulsion the instrument holds the valve and raises an alarm rather than switching. It will stop and wait for an operator before it lets a rag layer pass as either phase. The emulsion itself is a separation problem; what the instrument can do is refuse to wave it through.

Our line is DN32 and PTFE-lined — can this be fitted, and what happens at the liner?

Yes. DN32 is one of the standard bores (DN25 / DN32 / DN50). The lined build is a steel shell with a PTFE/PFA liner and sapphire windows that pass through the liner and wet the process.

Measuring through the liner is not an option: PTFE is 2.97 MRayl and strongly attenuating, and the bond between liner and shell changes with temperature and permeation — that turns a fixed acoustic path into a drifting one.

Two ordinary lined-pipe rules also apply. Wetted seals are PTFE or FFKM; ordinary FKM rates only fair against methylene chloride. And the shell's vent holes must stay clear — chlorinated solvents permeate fluoropolymer liners slowly, and trapped permeate lifts the liner off the wall.

Against conductivity and photoelectric (sight-glass) detectors — when should each be chosen?

Choose by whether the premise holds; each method rests on one.

Conductivity assumes the aqueous phase conducts, the organic does not, and an electrode may wet the process. Where that holds it is the cheapest answer. Pure-water washes, wet organics and lined pipe that is awkward to tap all soften it.

Photoelectric / sight glass assumes an optical difference. Turbidity- and colour-based units fail outright on dichloromethane and water — both are clear and colourless — while a refractive-index unit does work (1.424 against 1.333). Both share one weakness: a fouled window.

Sound velocity (PS7021) rests on two conditions only: the phases differ by at least 30 m/s, and the pipe runs full. It is independent of colour, turbidity, conductivity and density, which is why it holds where both phases are clear, both are non-conducting, or the aqueous conductivity moves. The price is a two-phase calibration on site, repeated at product changeover — five minutes, but somebody has to do it.

The item-by-item comparison is here.

Does commissioning need a shutdown, and how are the samples taken?

No shutdown. Calibration needs two velocity bands, and both samples come out of a normal batch: one while the light phase is running steady, one while the heavy phase is, 200 mL or more each. The instrument reads them itself — nothing goes to a lab. It usually fits inside one batch.

Three cautions: sample at process temperature (velocity is temperature sensitive, organics run −3 to −4 m/s per °C); draw the sample near the measuring point; and take the aqueous phase at its real salinity — clean water is not a substitute.

What you end up with is a velocity-versus-temperature table for both phases, delivered with the instrument. If the process temperature changes later, the thresholds shift along that table without a service visit.

How do the three relays drive a diverter valve, and which way does it fail?

Three volt-free contacts carry heavy, emulsion and light. The usual wiring: heavy closed → diverter to the heavy receiver; light closed → light receiver; emulsion closed → hold the current position and alarm for an operator to confirm.

The failure direction is the first thing to settle. Gas, flashing and window film all push the instrument toward not-decidable rather than toward the other phase, so making not-decidable mean stay put runs with the grain of the disturbances. Defaulting it to either phase turns every disturbance into a carry-over incident.

Keep the manual mode and the existing sight glass: the instrument supplies the judgement, not the valve. 4-20 mA trends velocity in the DCS; RS485 exposes phase, velocity, amplitude and temperature for the batch record.

How is it accepted on site — what do you compare it against?

The output is a phase decision, not a concentration you can compare against a titration, so acceptance looks different from a concentration meter's. Three workable tests:

1. Timing against the operator. Run a number of batches and align the instrument's switch instant with the operator's call at the sight glass. The site sets the tolerance; what usually matters is whether the instrument is ever late, since late means carry-over.

2. Sampling around the cut. Draw a sample either side of the instrument's switch instant and check that carry-over sits inside process limits.

3. The velocity itself. Measured band values should match the calibration table, and batch-to-batch drift in one phase should stay within the repeatability seen at calibration (±0.2 m/s). This is the test that separates "the instrument drifted" from "the material changed".

Can it measure how thick the rag layer is, or where the interface sits in a tank?

No — those are different instruments. PS7021 is an inline spool and knows only what is crossing its acoustic path right now: heavy, emulsion or light.

Interface height in a tank or settler belongs to an interface level instrument (magnetostrictive, radar, ultrasonic level, capacitance), mounted on the vessel rather than in the line. Rag layer thickness is the same question.

There is a useful indirect answer, though: time the instrument spends in the emulsion state during a drain, multiplied by flow, gives the volume of rag in that batch. It is not an interface height, but as a trend for how well the break worked or whether settling time is enough, it is often the more direct number.

Does it need a straight run? Can it go right after a pump or a control valve?

No straight run is required. The path length is fixed by the spool and the discriminators are velocity and amplitude, which do not care about the flow profile — unlike DP, vortex or magnetic meters, which live on it.

The position still matters, but for a different reason: phase and gas. Pump discharges, the downstream side of control valves and other throttling points generate gas or flashing continuously, which holds the amplitude down and degrades three-state discrimination. Mount upstream of them, or further downstream where the flow has settled.

Three other rules: a full line is not negotiable; on horizontal runs put the path across the horizontal (3 and 9 o'clock), away from gas at the top and solids at the bottom; and keep the measuring point close to the diverter, because that hold-up is the mixed volume every batch pays.

What about high temperature, sanitary execution and CIP?

Temperature: −20 to +80 °C standard, −20 to +120 °C for the high-temperature variant. Heat actually helps the discrimination: organic velocity falls with temperature while water rises below 74 °C, so the bands spread. Edible oil washing at 80-90 °C is the textbook case.

Sanitary: the standard spool is industrial and flanged. Tri-clamp, polished bore and drainable geometry are a custom build with their own lead time and price — say so at selection rather than assuming the standard spool can enter a sanitary zone.

CIP: the sapphire window is dense and smooth, so hot caustic, hot water or solvent all clean it in place. Compare the amplitude baseline before and after and you know whether the clean worked — far easier than scheduled teardown.

Can it still work if the line carries solids or crystals?

It depends on the load. A light suspension does not move the velocity reading but scatters sound and lowers the echo amplitude — and amplitude is what catches the rag layer, so with heavy solids three-state discrimination degrades to two: light and heavy still separate, emulsion and "solids present" no longer do.

In practice: put the point after settling, filtration or a steady pump; at commissioning re-baseline the amplitude against the real solids-bearing fluid rather than against clear liquid; and if solids content is itself the quantity you want, that is another instrument — PS7000 and the acoustic-impedance method.

Crystals add one more wrinkle: they grow a film on the window. The film shows in the amplitude baseline first, which makes that baseline a useful clean-me reminder.

How wide does the velocity gap have to be, and can it be estimated first?

The floor is 30 m/s at working temperature. Common organic-against-water pairs are far above it: carbon tetrachloride is 572 m/s from water, dichloromethane 428, ethyl acetate 413, toluene 170, m-xylene 155 (handbook values at 25 °C). For those the gap is never the constraint.

Three cases do need arithmetic first: oil against oil (diesel and kerosene are 74 m/s apart), oil against water at ambient (edible oil and water, about 66), and formulated fluids — peroxide working solution, SX organic, a demulsified waste oil.

How to estimate: take handbook values at 25 °C, carry both to working temperature with their coefficients (organics −2.7 to −4.9 m/s per °C, water about +2.4 below 74 °C), then add salt on the aqueous side (roughly +11 m/s per wt% NaCl). Above 100 m/s you are safe; 30 to 100 needs a measurement; below 30, don't.

How many products can one instrument cover, and does a changeover need recalibration?

Eight recipes as standard, each holding a pair of velocity bands and its amplitude threshold; the list extends. Switch from the panel, or let the DCS write a Modbus register — the second follows the batch instruction and removes the "nobody switched it" class of incident.

A product that has never been calibrated needs one pass: two samples, two readings, one stored pair — five minutes, no return-to-factory sampling. Coming back to a known product just recalls its recipe.

When to redo an existing recipe: the composition of that phase moved (water content, salt, product concentration), the temperature band shifted, or acceptance found measured velocity off the stored band by more than the repeatability. Every recalibration records its time and both velocities for the batch record.

How often does the sapphire window need cleaning, and how do you know?

No fixed interval — it depends on the duty. What actually helps: the instrument tells you.

As film, crystal or oil builds on the window, the velocity reading does not move at first, but the amplitude baseline drifts down. Trend the clear-phase baseline: a slow decline means clean it; a sharp drop with velocity still in band is an emulsion slug passing, not deposit. The two shapes are easy to tell apart on a trend.

How to clean: sapphire is dense and smooth, so hot caustic, hot water or a normal solvent will do it, usually in place. On duties that will certainly film — edible oil, spent emulsions — put it on the CIP schedule. Afterwards check whether the amplitude baseline came back: that is the acceptance test for the clean.